Int Poster J Dent Oral Med 11 (2009), No. 1     15. Mar. 2009

Int Poster J Dent Oral Med 2009, Vol 11 No 1, Poster 434

Clinical study of postoperative sensitivity for a new self-adhesive resin cement

Language: English
 

Authors:
Dr. Constanze Olms, Dr. Arne Boeckler, Prof. Dr. Jürgen-Manfred Setz,
Department of Prosthodontics,
Dr. Christine Lautenschläger,
Institute of Medicinal Epidemiology, Martin-Luther-University Halle-Wittenberg

Date/Event/Venue:
July 2-5, 2008
86th General Session & Exhibition of the IADR
Toronto
 

Introduction

Self-adhesive resin cements have gained popularity over the past few years because of their improved physical properties as compared to conventional cements. New luting agents have simplified clinical handling, a wider range of applications and reduced pulp injury. Postoperative hypersensibility and pulp injury was a problem in the past [2]. The literature shows data for postoperative hypersensibilities between 0% - 19% [3-5]. With regard to conventional adhesive luting agents a new self-adhesive resin cement provides simplified clinical procedure. Under the condition of relative dryness no additional dentin bonding is necessary. The dentin-restoration-bonding is realised by means of a phosphoric acid containing monomer. So, the tested cement is applied directly to the dentin. In-vitro tests proofed the quality of the cementation to dental alloys, zirconium oxide and glass ceramics as well as fibre enforced endodontical posts.[1].
 

Objectives

The aim of this clinical study was to evaluate a new self-adhesive resin cement for possible postoperative sensitivity. Additionally changes in cement colour, quality of the marginal seal, as well as the occurrence of de-cementations were registered.
 

Material and Methods

Nineteen patients received 60 full-coverage restorations on 43 vital and 17 successfully endodontically treated abutment teeth in the time period between May 2006 to December 2007. The crowns were cemented with a new self-adhesive resin cement (Multilink® Sprint, Ivoclar Vivadent, Schaan Liechtenstein) according to the manufacturers' instructions (Fig. 2-11). Documentation began with the cementation of the crowns. The teeth were inspected immediately after cementation, after 2 weeks (baseline), 6 months and 18 months. Tooth vitality was examined with an -50° C ice spray (roeko Endo-Frost, Fa. Coltène Whaledent, Langenau Germany) before and after cementation, after 2 weeks, 6 months and 18 months. The prevalence of follow-up postoperative sensitivities compared to sensitivity after cementation was analyzed (Tab. 1). The statistically significance was proofed by means of Chi2 tests.

Fig. 1
Fig. 2 Fig. 3
Fig. 4 Fig. 5
Fig. 6 Fig. 7
Fig. 8 Fig. 9
Fig. 10 Fig. 11
 

Results

After cementation postoperative hypersensitivity was diagnosed in 7% of the abutment teeth (Fig. 1, Tab. 1). At baseline (after 2 weeks) the rate was reduced to 4.7%. Postoperative hypersensitivity was not reported after 6 months (Tab. 2). One patient with two adjacent crowns showed hypersensitivity after 18 months. This was attributed to the fact that the gingiva had retracted and a hypersensitive dentin area had been uncovered. In one case endodontic treatment was carried out for one abutment tooth after 12 months. Cement colour and marginal seal did not change. De-cementation did not occur during the period of observation.

p – 0.62

  postoperative hypersensitivity
after 2 weeks
 
normal hypersensitive total
postoperative hypersensitivity
after cementation
normal number 38 1 39
% totality 88.4% 2.3% 90.7%
hypersensitive number 3 1 4
% totality 7.0% 2.3% 9.3%
total   number 41 2 43
% totality 95.3% 4.7% 100.0%
Tab. 1
 
  postoperative hypersensitivity
after 6 month
 
normal total
postoperative hypersensitivity
after cementation
normal number 37 37
% totality 90.2% 90.2%
hypersensitive number 4 4
% totality 9.8% 9.8%
total   number 41 41
% totality 100.0% 100.0%
Tab. 2
 

Conclusions

Compared to existing literature for dental cements the new self-adhesive cement showed a small rate of postoperative hypersensitivities [2-5]. The overall clinical experience proofed the cement to be an encouraging alternative to conventional cements with regard to physically data, postoperative sensitivity and ease of application. Additional prospective trials remain necessary to verify the clinical long-term performance of the tested product.
 

Literature

  1. Christensen GJ. Resin cements and postoperative sensitivity. J Am Dent Assoc 2000, 131, p. 1197-1199
  2. Denner N, Heydecke G, Gerds T, Strub JR. Clinical comparison of postoperative sensitivity for an adhesive resin cement containing 4-META and a conventional glass-ionomer cement. Int J Prosthodont 2007, 20, p. 73-78
  3. Jokstad A. A split-mouth randomized clinical trial of single crowns retained with resin-modified glass-ionomer and zink phosphat luting cements. Int J Prosthodont 2004,17, p. 411-416
  4. Johnson GH, Powell LV, DeRouen TA. Evaluation and control of post-cementation pulpal sensitivity. Zinc phosphate and glass ionomer luting cements. J Am Dent Assoc 1993, 124, p. 38-46
     

This Poster was submitted by Dr. med. dent. Constanze Olms.
 

Correspondence address:
Dr. med. dent. Constanze Olms
Martin-Luther-University Halle-Wittenberg
Department of Prosthodontics
Große Steinstraße 19
06108 Halle
Germany